:STATus: OPERation?
Read Operation Status, Event
Reads out the contents of the operation event status register. Reading the Opera
-
tion Event Register clears the register. See figure on page 8-107.
Returned Format: <Decimal data>¿
<dec.data> = the sum (between 0 and 368) of all bits that are true. See table on page 8-108.
Complies with standards: SCPI 1991.0, confirmed.
:STATus :PRESet
Enable Device Status Reporting
This command has an SCPI standardized effect on the status data structures. The
purpose is to precondition these toward reporting only device-dependent status
data.
It only affects enable registers. It does not change event and condition registers.
The IEEE-488.2 enable registers, which are handled with the common commands *SRE and
*ESE remain unchanged.
The command sets or clears all other enable registers. Those relevant for this counter are as
follows:
It sets all bits of the Device status Enable Registers to 1.
It sets all bits of the Questionable Data Status Enable Registers and the Operation Status En
-
able Registers to 0.
The following registers never change in the counter, but they do conform to the standard
:STATus:PRESet values.
All bits in the positive transition filters of Questionable Data and Operation status registers
are 1.
All bits in the negative transition filters of Questionable Data and Operation status registers
are 0.

Table of Contents

Other manuals for Pendulum CNT-90

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pendulum CNT-90 and is the answer not in the manual?

Pendulum CNT-90 Specifications

General IconGeneral
BrandPendulum
ModelCNT-90
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals