EasyManuals Logo

Pentair HAFFMANS c-DGM User Manual

Pentair HAFFMANS c-DGM
79 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #8 background imageLoading...
Page #8 background image
HAFFMANS c-DGM/i-DGM
Page 8 of 78 c-DGM/i-DGM
* at 20 °C : *O
2
dissolved
LHO (DO)
:[ppb w/w]
: ± (1 ppb+2%
of the measured value)
: [µg/l] : ± (1 µg/l +2% otmv)
: [ppm w/w] : ± (0,001 ppm+2% otmv)
: [mg/l] : ± (0,001 mg/l+2% otmv)
: *O
2
in gas : [%a.s.] : ± (0,01%a.s.+2% otmv)
: [%] O
2
: ± (0,002 %O
2
+2% otmv)
: [ppm
vol/vol]
: ± (20 ppm+2% otmv)
:*O2 dissolved
WLO (DO)
: [ppb w/w]
:0 – 45000 (<5% otmv
from 100 ppb)
: [µg/l]
:0 – 45000 (< 5% otmv
from 100 µg/l)
: [ppm w/w]
:0 – 45 (<5% otmv from
0.1 ppm)
: [mg/l]
:0 – 45 (<5% otmv from
0.1 mg/l)
: *O2 in gas :[%a.s.]
: 0 – 500 (<5% otmv from
1.098 %A.s)
:[%] O
2
: 0 – 100 (<5% otmv from
0.230 %O
2
)
: [ppm
vol/vol]
: 0 – 1000000 (<5% otmv
from 2300 ppm)
: Temperature : [
o
C]
: [
o
F
: ± 0.2
: ± 0.4
: Pressure : [barg] : ± 0.01
: [psi] : ± 0.2
: [kg/cm
2
] : ± 0.01
: [kPa]
: ± 1
Maximum pressure during
operation
: [barg]
: [psi]
: [kg/cm
2
]
: [kPa]
: 10.00
: 145.0
: 10.20
: 1.000
Maximum cleaning
temperature (short duration)
: [
o
C]
: [
o
F]
: 60
: 140
Minimum conductivity (CO2
2
measurement)
: [µS/cm] 400

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pentair HAFFMANS c-DGM and is the answer not in the manual?

Pentair HAFFMANS c-DGM Specifications

General IconGeneral
BrandPentair
ModelHAFFMANS c-DGM
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals