EasyManua.ls Logo

PerkinElmer DSC 6000

PerkinElmer DSC 6000
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Install a DSC 6000 . 43
The PCD (P/N N5202068) can be filled with ice, ice water, liquid nitrogen or any other
cooling mixtures.
The DSC 6000 with CryoFill uses liquid nitrogen for operation in the range of 170 °C to
300 °C.
NOTE: When operating the analyzer with a cooling device you MUST use a purge gas.
Connecting to the Chiller
When using the chiller, connect the cooling supply to the DSC 6000 as follows:
CAUTION
ATTENTION
An appropriate fluid must be used with the chiller. The fluid should be
effective down to 20 °C. Any lower than that would impair the
conveyance power of the chiller's pump. Distilled water is recommended
for operation at temperatures between 15 °C and 90 °C. For operation at
20 °C to 100 °C, 50% ethylene glycol in water is recommended.
Un fluide approprié doit être utilisé avec le refroidisseur. Le fluide devrait
être efficace jusqu'à -
20 ° C. Tout inférieur à celui qui affecterait le
pouvoir de transport de la pompe du refroidisseur. L'eau distillée est
recommandée pour un fonctionnement à des températures comprises
entre 15 ° C et 90 ° C. Pour une utilisation de -20 ° C à 10
0 ° C, 50%
d'éthylène glycol dans l'eau sont recommandés.
CAUTION
ATTENTION
Ensure that the cooling air vents at the front and rear of the cooler are
not blocked. Maintain a minimum space of 1520 cm.
Assurez-vous que les évents d'air de
refroidissement à l'avant et à
l'arrière du refroidisseur ne sont pas bloqués. Maintenir un espace
minimum de 15-20 cm.
1. Make sure that the chiller is switched off but that its power cord is plugged into a power
supply.
2. Screw one of the barbed hose couplings shipped with the chiller into the OUTLET
connection on the chiller unit, and tighten with a suitable wrench.
3. Couplings suitable for several different hose sizes are provided with the chiller.
4. Push a hose (compatible with the desired coolant) onto the OUTLET coupling and
secure with a hose clamp.
5. Slide the other end of the hose over the metal tube labeled COOLING LIQUID IN on the
thermal analyzer. Place a hose clamp around the hose and metal tube to secure the
hose in place.
6. Screw a barbed hose coupling into the INLET connection on the chiller unit, and tighten
with a suitable wrench.
7. Push a hose onto the INLET coupling and secure with a hose clamp.

Table of Contents

Related product manuals