EasyManua.ls Logo

Pfeiffer Vacuum ASM 340 - Commissioning; Detector start-up; Detector powering off

Pfeiffer Vacuum ASM 340
108 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Commissioning
6.1 Detector start-up
1. Connect the main power supply cable.
2. Set the main switch/circuit breaker to I.
3. For first start-up: set the language, unit, date and time (the user can modify these settings at a
later time).
4. Wait for the detector to enter “Stand-by” mode.
Start-up after an extended shutdown
If the detector has been stored or has not been used, start-up time is longer than if it is in regular use.
After a shutdown of the detector of more than 3 months, it is advisable to start up the detec-
tor 24 hours before its use.
6.2 Test launched automatically upon start-up
This function is used to automatically start the first test after the detector start-up phase.
If the function is enabled, the 1
st
test starts automatically as soon as the detector start-up phase is com-
pleted. Subsequent tests are launched by the user.
1
1 Check box for the ‘Run Test after start?’ message
1. Ensure that the inlet is connected before starting the function.
2.
During the detector start-up phase, check [] to enable the function.
The question will be asked each time the leak detector is switched on. The last selection
made is not saved.
6.3 Detector powering off
1. Set the switch/circuit breaker to O.
2. Wait for the control panel screen to turn off completely before working on the product, removing
the cover and/or moving it.
3. Disconnect the mains power cable.
Pump stop due to power failure
When there is a mains power failure, the detector shuts down: it switches on again automatically when
power is restored.
Commissioning
31/108

Table of Contents

Other manuals for Pfeiffer Vacuum ASM 340

Related product manuals