Maintenance
71
7c With the J shaped tube and hex union still attached to the capillary, pull the
capillary out of the heater through the front oven. See Figure 10-8.
Figure 10-8 Remove stainless steel capillary assembly
7d Disconnect the hex union from the old capillary.
7e Install a nut and ferrule (B0141949) on one end of the new capillary and place
them in the open end of hex union.
7f Tighten the hex nut using 1/4" wrenches.
7g Insert the free end of the capillary through the oven opening and into the
opening of the capillary heater. Push the entire length of capillary through the
heater. See Figure 10-9.
7h Connect the J shaped tube to the inlet and tighten the nut using the nut driver.
Figure 10-9 Attach stainless steel capillary assembly
NOTE
Ensure that the shape of J shaped capillary is not modified.
Hex
Union
J Shaped
Tube