EasyManua.ls Logo

Pfeiffer TPG 261 - Page 29

Pfeiffer TPG 261
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BG 5195 BEN / A (2011-05) TPG261.oi 29
Pressure p
Measurement
range
If the TPG 261 is operated with linear gauges
(CMR 261 … 375, APR 250 … 267), negative pressures
may be indicated.
Possible causes:
negative drift
offset correction activated with positive offset.
Measurement range

Related product manuals