BG 5195 BEN / A (2011-05) TPG261.oi 83
Transmit: UNI [,x] <CR>[<LF>]
└─ Pressure unit, x =
0 –> mbar/bar (default)
1 –> Torr
2 –> Pascal
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x <CR><LF>
└─ Pressure unit
Transmit: BAU [,x] <CR>[<LF>]
└─ Transmission rate, x =
0 –> 9600 baud (default)
1 –> 19200 baud
2 –> 38400 baud
As soon as the new baud rate has been en-
tered, the report signal is transmitted at the new
transmission rate.
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x <CR><LF>
└─ Transmission rate
Transmit: DCD [,x] <CR>[<LF>]
└─ Resolution, x =
2 –> Display x.x (2 digits)
(default)
3 –> Display x.xx (3 digits)
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x <CR><LF>
└─ Resolution
When the PrE ( 46) is ON and the pressure is in the
range p<1.0E-4 mbar the display resolution of the PCR
Gauge is reduced by one decimal digit.