EasyManua.ls Logo

Pfeiffer TPG 261 - Page 49

Pfeiffer TPG 261
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BG 5195 BEN / A (2011-05) TPG261.oi 49
Certain gauges can be activated by different means.
The following gauges can be controlled:
Pirani Gauge
(TPR)
Pirani Capacitance Gauge (PCR)
; Cold Cathode Gauge
(IKR)
; FullRange™ CC Gauge
(PKR)
; Process Ion Gauge
(IMR)
; FullRange™ BA Gauge
(PBR)
Capacitance Gauge
(CMR)
Piezo Gauge
(APR)
Value
Ö Manual activation:
The gauge can be activated by
pressing the
key.
Ö External activation:
The gauge is activated by an
input signal fed via the control
connector ( 21).
Ö Hotstart:
The gauge is automatically ac-
tivated when the TPG 261 is
turned on. Measurement is
thus automatically resumed
after a power failure. Gauge
deactivation 50.
Ö Increase/decrease the value
by the defined increments.
Gauge activation

Related product manuals