13
FREEZER COMPONENTS
1. Control panel: The chamber temperature and other alarms/functions can be set through the keys on
the control panel. The operational status can be checked on the temperature display and indicators
[page 14].
2. Lock: By using the enclosed key, both the upper door and lower door can be locked or unlocked.
3. Levelling foot: These are screw bolts used to install and fix the unit. Adjust the height of the levelling
feet by turning the screw bolts until the two front casters are away from the floor [page 16].
4. Space for temperature recorder: A temperature recorder (optional) can be mounted here so that the
chamber temperature can be recorded automatically [page 37].
Contact our sales representative or agent for the installation.
5. Storage container: These are made of styrol resin. Be careful not to damage the storage containers
when removing frost with the scraper.
The bottom 2 large storage containers are used for collecting the condensate water during defrosting
[page 29].
Note:
Be careful when pulling the storage containers to ensure they do not fall.
6. Door latch: Always lock the door latch when the door is closed.
7. Fixture: Use the fixtures to secure the unit to a wall with a strong rope or chain [page 16].
8. Space for optional component: An interface board (MTR-480) or a LAN interface board (MTR-L03)
can be installed here.
9. Remote alarm terminal: A remote alarm device (separately available) can be connected to this
terminal. The remote alarm relays the alarm to an operator in a remote location if the unit is un attended.
Refer to page 25 for details.
Contact our sales representative or agent to arrange the installation.
10. Battery switch: This is an ON-OFF switch for the battery for the power-failure alarm. Always turn
this switch on when the unit is operating to ensure that the power-failure alarm is working. Turn this
switch off when the unit is not used for a long period in order to protect the battery.
11. Power switch: This is a power switch of the unit (ON – “I” OFF – “
○”).
12. Access port: This port is used to pass the sensor or cable of measuring equipment, or sensor of
temperature recorder (optional) to the chamber.
Replace the insulation and the rubber caps when the access port is not in use. Improper replacement
may cause an increase in chamber temperature or condensation around the access port.