SETTING THE ALARM TEMPERATURE
Setting the low-temperature alarm
An abnormality (chamber temperature drop) is indicated by the alarm indicator, and the chamber
temperature display blinking, and the buzzer sounding (15 minutes after blinking) if the chamber
temperature falls below the set value of low-temperature alarm. Always set the low-temperature alarm to
protect stored items from damage resulting from a drop in temperature.
Setting range of work temperature for low temperature alarm:
Between 5
o
C and 15
o
C lower than the chamber set temperature
Initial setting (factory setting): 10
o
C lower than the chamber set temperature
For example, to change the alarm temperature from chamber set temperature minus 10
o
C (i.e. the
low-temperature alarm works at 10
o
C lower than the chamber set temperature) to chamber set
temperature minus 5
o
C (i.e. the high-temperature alarm works at 5
o
C lower than the chamber set
temperature)
(The following is only an example. Choose the actual setting according to the desired alarm
temperature.)
Operation Key operated Display after operation
1
-----
The current chamber temperature is
displayed.
2
Press the up arrow key for 5
seconds.
The current chamber temperature
changes to function code F00 and
the right-hand figure (0) blinks.
3 Press the up arrow key twice.
The display changes to from F00
to F02.
4 Press the “Set” key once.
SET
The current setting (-10) is displayed
and the right-hand figure (0) blinks.
5
Press the up arrow key (as many
times as needed to reach the
desired figure).
The display changes from -10 to -15.
6 Press the scroll key once.
The middle figure (1) starts to blink.
7
Press the up arrow key (as many
times as needed to reach the
desired figure).
The display changes from -15 to -05.
8 Press the “Set” key once. SET
The new setting of the low-temperature
alarm is memorized and the current
chamber temperature is displayed.
The unit will return from setting mode to temperature display mode automatically after 90 seconds if no
key is operated (auto-return function). In this case, the setting is not changed.