EasyManua.ls Logo

Philips Optimus - 15.1.1.1 Connecting and Setting the Scope

Philips Optimus
53 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
During alignment this factor duty cycle must be entered via AGenT. The influence of this factor as a correction
value for the Z-data table is monitored as the
U
COMP
signal, since the PI-controller is deactivated by the grounded
GAIN_IN
signal.
15.1.1.1 Connecting and setting the scope
For connections see figure below:
Status
EZ130
GNDA
GAIN_IN
AV CA
AV AN
AV HT
UCOMP
Generator front
EZ130 front
Osci: Y1 ==> X3
X4
X5
X6
X23
Osci: Y2 ==> X26
X32
H800
X800
Short link
Ground the probes in
one of these drilling holes
(Fig. 27)
Channel 1 = EZ130 X3 ---> AV HT ---> 20kV/V ---> 1V/div --> Zero-line at bottom of screen
Probe GND = one of the drilling holes at the front cabinet chassis
Channel 2 = EZ130 X26 ---> UCOMP ---> 1V/div ---> Zero-line 2 div from bottom of screen
Probe GND = one of the drilling holes at the front cabinet chassis
Trigger = external (preferred) ---> CTRL_X_C/ ---> backpanel EZX74 / negative slope
or = internal channel 1 ---> AV HT ---> EZ130 X3 / positive slope at +3V
Probe GND = one of the drilling holes at the front cabinet chassis
Time base = 5 or 10ms/div ---> trigger delay -1div
NOTE
A digital scope should not have any other ground connection than the ground of the
three probes at the drilling holes at the front generator chassis.
A mains-driven scope must be isolated from earth potential, otherwise it might display
artefacts.
Converter test kit OPTIMUS for OPTIMUS 50/65/80 gen‐
erators release 3.x with converters 4512 104 7231x
42 CSIP Level 1 (08.0)
© 2008 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Conv test Optimus

Table of Contents

Related product manuals