10 TUBE CHECK WITH ROTOR CONTROL PART 2
Switch ON the generator.
The system and the generator should be in ready condition.
Check of the tube vacuum:
Push and hold PREP.
Switch OFF the generator.
The anode is in free run now.
Listen at the tube housing:
If the anode speed is down within 2 ... 3 minutes the tube is defective (no vacuum) and must be replaced.
If a decrease of anode speed is not audible within several minutes the vacuum is still OK.
Switch the generator back ON.
A generator with high-speed rotor control brakes the tube automatically (chapter 8).
At generators with low -speed rotor control wait for ready condition, push PREP once and let go of the
PREP switch to brake the tube.
Converter test kit OPTIMUS for OPTIMUS 50/65/80 gen‐
erators release 3.x with converters 4512 104 7231x