14 FINAL SYSTEM TEST
▪ If the generator works error-free at this stage switch it OFF and cut it from mains.
WARNING
Wait at least 1 minute until the converter DC is discharged.
▪ Remove current transformer(s) from the primary lines and reestablish the primary lines at the resonance ca‐
pacitors.
▪ Close the converter covers.
▪ Put mains and the generator back ON.
▪ Run a final system test with exposures and fluoroscopy, where present.
Converter test kit OPTIMUS for OPTIMUS 50/65/80 gen‐
erators release 3.x with converters 4512 104 7231x