Defining Setup Options
8-8 LIFEPAK 20 Defibrillator/Monitor Operating Instructions
PACING SETUP MENU
The Pacing Setup menu allows you to define noninvasive pacemaker settings. When you select a
menu item, the screen displays a help message. The options in bold are factory default settings.
MONITORING MENU
Use the Monitoring menu to define settings for the ECG and SpO2 monitoring. When you select a
menu item, the screen displays a help message. The options in bold are factory default settings.
Channels Setup Menu
To define the default set and up to five optional waveform sets for Channels 1 and 2, select an item
from the Channels Setup menu.
Table 8-9 Pacing Setup Menu
Menu Item Help Message Options
RATE
Default pacing rate 40ā170, 60.
CURRENT
Default pacing current 0ā200 mA.
MODE
Default pacing mode DEMAND or NONDEMAND.
INTERNAL PACER
Detect internal pacemaker and
print arrows
DETECTION OFF or DETECTION ON.
Table 8-10 Monitoring Menu
Menu Item Help Message Options
CHANNELS...
Set up default channel waveforms Refer to Ta bl e 8 -1 1 .
CONTINUOUS ECG
Continuously store ECG
waveform
ON or OFF.
SPO2 TONE
SpO2 pulse tone ON or OFF.
CO2...
Set up CO2 defaults Refer to Ta bl e 8 -13 .
Table 8-11 Channels Setup Menu
Menu Item Help Message Options
Default SET Select default waveform set Set 1, Set 2, Set 3, Set 4, or Set 5.
SET 1
Select channel waveforms for
Set 1
Refer to Ta ble 8 -1 2.
SET 2
Select channel waveforms for
Set 2
Refer to Ta ble 8 -1 2.
SET 3
Select channel waveforms for
Set 3
Refer to Ta ble 8 -1 2.
SET 4
Select channel waveforms for
Set 4
Refer to Ta ble 8 -1 2.
SET 5
Select channel waveforms for
Set 5
Refer to Ta ble 8 -1 2.