EasyManua.ls Logo

Physio Control LIFEPAK 20 - Replacing and Removing Electrodes

Physio Control LIFEPAK 20
202 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 Paddle Accessory Options
Paddle Accessory Options
LIFEPAK 20 Defibrillator/Monitor Operating Instructions 5-5
©2002–2013 Physio-Control, Inc.
For adult patients, follow the procedures for ECG monitoring, AED defibrillation, manual defibrillation,
synchronized cardioversion, and pacing described in Section 3 or Section 4.
For pediatric patients, follow the procedures for ECG monitoring, manual defibrillation, synchronized
cardioversion, and pacing except for the following:
• Select the appropriate defibrillation energy for the weight of the pediatric patient according to the
American Heart Association (AHA) recommendations or local protocol. Using energy levels of
100 J or greater is likely to cause burns.
• When pacing, frequently inspect the patient’s skin under the heart electrode for signs of burns.
Note: The amount of pacing current needed for capture is similar to the pacing current needed
for adults.
Replacing and Removing Electrodes
Replace QUIK-COMBO, QUIK-COMBO RTS, QUIK-COMBO REDI-PAK, or FAST-PATCH Plus
electrodes after 50 defibrillation shocks or 24 hours on the patient’s skin, or after 8 hours of
continuous pacing. Replace pediatric QUIK-COMBO electrodes after 25 defibrillation shocks or
24 hours on the patient’s skin, or after 8 hours of continuous pacing.
To remove QUIK-COMBO or FAST-PATCH electrodes from the patient:
1 Slowly peel back the electrode from the edge, supporting the skin as shown in Figure 5-5.
Figure 5-5 Removing Therapy Electrodes from Skin
2 Clean and dry the patient’s skin.
3 When replacing electrodes, adjust the electrode positions slightly to help prevent skin burns.
4 Close the protective cover on the QUIK-COMBO therapy cable connector when the cable is not in
use.
To disconnect the defibrillation cable from the FAST-PATCH electrodes:
1 Press down around the electrode post.
2 Pinch the snap connector with the fingers of the other hand and pull straight up (refer to
Figure 5-6).
Figure 5-6 Disconnecting Defibrillation Cable from FAST-PATCH Electrodes

Table of Contents