4Therapy
Therapy
LIFEPAK 20 Defibrillator/Monitor Operating Instructions 4-9
©2002–2013 Physio-Control, Inc.
Proper Connection
If therapy electrodes are not connected to
the therapy cable or not placed on the
patient’s chest, the
CONNECT ELECTRODES
message and voice prompt occur until the
patient is connected to the AED.
If the therapy cable is not connected to the
defibrillator, you will see the
CONNECT
CABLE
message until the cable is
connected.
If the test plug is connected to the therapy
cable when analysis is initiated, the
REMOVE
TEST PLUG
message and voice prompt
occurs.
After you remove the test plug from the
therapy cable, the message and voice
prompt
CONNECT ELECTRODES occurs until
the patient is connected to the AED.