EasyManua.ls Logo

Picotest M3500A - Simple Calibration Procedures; The Calibration Security Code; The Zero Adjustment

Picotest M3500A
94 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
85
6 Simple Calibration Procedures
If users’ multimeters need detailed calibration service, contact Picotest Corp
for more information. If users only want to do the basic calibration for Zero
& Gain adjustments, please follow the procedures below.
6.1 The Calibration Security Code
To prevent from unauthorized or abusive calibrations, generally, the
M3500A is secured by a specific code. As an M3500A needs calibration,
operators have to input a security code. Then, the M3500A would just enter
the calibration mode. The decoding procedures are in the following.
MENU > PREV > CAL MENU > ENTER > SECURED > ENTER > CODE > ENTER
> input 123456 (by using the UP button in the Range area & NEXT button) >
ENTER
Users can execute the calibration/adjustment procedures by entering
UNSECURED > NEXT > CALIBRATE > ENTER.
Note 1: The factory default is 123456, which is burned in the
permanent memory of the M3500A. If the default needs to be defined
for purposes, please use PT-SET software.
6.2 The Zero Adjustment
Please follow the procedures A, B and C to do the Zero adjustment.
A. Use the CALIBRATE SHORT PIN as shown below to plug into the front
panel terminals of the M3500A. Then, warm it up in 3 minutes as the
M3500A is given power.
B. After lifting the security, please do the Zero adjustment by the operation
procedures, such as UNSECURED > NEXT > CALIBRATE > ENTER (The
display will show input full scale”)> input +0.000000E-1 (by using the

Other manuals for Picotest M3500A

Related product manuals