88
Table-1
6.5 Calibration Count
The calibration count will be increased from finishing calibration each time.
According to the calibration procedures on various points, a full calibration
will generate the count number more than one, and it will be recorded in the
permanent memory of the M3500A.
Note 2: The factory default for each M3500A is 0. If the count number
exceeds 32767. The number will turn into 0.
6.6 Calibration Date
The M3500A will record a calibrated and a future calibration dates as the
factory default. Please note that the dates are not allowed to be revised.
6.7 The Calibration Error Messages
Please take a look at the following messages (Table-2) as calibration errors
occur.
Table-2
ACV
100mV to 100V
0.9 to 1.1 * 10% * full scale
0.9 to 1.1 * 100% * full
scale
0.9 to 1.1 * 10% * full scale
0.9 to 1.1 * 100% * full
scale
750V
0.9 to 1.1 * 10% * full scale
675V to 770V
0.9 to 1.1 * 10% * full scale
675V to 770V
FREQ Any
Any input > 100 mV
1K to 300K
Any input > 100 mV
1K to 300K