EasyManua.ls Logo

Pixsys ATR 401 - LATCH on Functions (Only AI1)

Pixsys ATR 401
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22 ATR401 - User manual
Updating Memory Card
To update the memory card values, follow the procedure described in the
rst method, setting display 2 to  so as not to load the parameters on
controller
1
.
Enter con guration level by password and change at least one parameter
(display will start  ashing). Exiting con guration mode, the settings will be
automatically saved on Memory card.
9 LATCH ON Functions (only AI1)
For use with input  (potentiometer 6 KΩ) and  (potentiometer
150 KΩ) and with linear input (0…10 V, 0...40 mV, 0/4…20 mA), it is possible
to associate start value of the scale (parameter 6  ) to the minimum
position of the sensor and value of the scale end (parameter 7 ) to the
maximum position of the sensor (parameter 8  con gured as ).
It is also possible to x the point in which the controller will display 0 (however
keeping the scale range between  and ) using the “virtual zero”
option by setting  or  in parameter 8 
If you set  the virtual zero will reset after each activation of the device;
if you set  the virtual zero remains  xed once tuned.
To use the LATCH ON function,  rst choose the selected option on parameter

2
Then refer to the following table for the calibration procedure:
Press Display Do
1
FNC
Exit parameters con guration.
Display 2 visualizes writing

Place the sensor on minimum
operating value (corresponding
to 
2
Store value on minimum.
Display shows

Place sensor on maximum
operating value (corresponding
to ).
3
Store value on max. Display
shows 
To exit standard proceeding
press
FNC
.
For “virtual zero” setting, place
the sensor to zero point.
1
If on activation the controller does not display  it means no data have been saved on the
memory card, but it is possible to update values.
2
The tuning procedure starts by exiting the con guration after changing the parameter.

Table of Contents

Related product manuals