CHECKING THE ALIGNMENTS
26 Planmeca ProOne
Installation Manual
9 CHECKING THE ALIGNMENTS
9.1 X-ray beam
The beam alignment must be checked after installation. Refer to the ProOne X-ray unit
Technical manual, section “Adjusting and calibrating X-ray beam”.
Adjust the X-ray beam according to the instructions given in the Technical manual, if
necessary.
9.2 Taking a ball phantom exposure
The position of the patient support table must be checked after installation. Take a ball
phantom exposure according to the instructions given in the Technical manual, section
“Taking a ball phantom exposure”.
Adjust the patient support table according to the instructions given in the Technical
manual, if necessary.
9.3 Patient positioning lights
Check the position of the patient positioning lights (midsagittal plane light, layer light and
Frankfort plane light) according to the instructions given in the Technical manual, section
“Adjusting the positioning lights”.
Adjust the patient positioning lights according to the instructions given in the Technical
manual, if necessary.
Printed copies of this document are considered uncontrolled.
7535.13.Rev001 08.07.2018