Ambient conditions during operation
Temperature °C +10 to +35
Relative humidity % 20 - 80
Air pressure hPa 750 - 1060
Height above sea level m < 2000
Ambient conditions during storage and transport
Temperature °C -18 to +60
Humidity % 10 - 95, non-condensing
Air pressure hPa 500 - 1060
Height above sea level m < 16000
Classification
Medical Devices Directive (93/42/EEC): Class I
Laser class (unit) in accordance with EN 60825-1:2014: 1
Laser source
Laser class 3B acc. to IEC 60825-1:2014
Wavelength nm 635
Power mW <10
Technical data of the RFID Module
Frequency MHz 13.56
Modulation ASK
Max. power mW 400
Electromagnetic compatibility (EMC) interference emission measurements
High-frequency emissions in accordance with
CISPR 11 Group 1
Group 1, Class B
Interference voltage at the power supply
connection CISPR 11:2009+A1:2010
Compliant
Electromagnetic interference radiation CISPR
11:2009+A1:2010
Compliant
Electromagnetic compatibility (EMC) interference immunity measurements
cover
Immunity to interference, discharge of static
electricity
IEC 61000-4-2:2008
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Compliant