4.4 Maintenance
4.4.1 Recommended maintenance schedule
Prior to working on the device or in case of danger, disconnect it from the
mains (e. g. pull the mains plug).
The recommended maintenance intervals are based on using the device for
25 intraoral images per day on 220 working days per year.
Maintenance interval Maintenance work
Annually Visually inspect the device.
Check the image plates for signs of scratches and change if necessary.
Remove dust and dirt from accessible parts.
Carry out a system check.
Every 3 years Exchange the cartridges.