Electromagnetic compatibility (EMC) Interference immunity measurements
supply input
Immunity to interference, rapid transient bursts –
AC voltage grid
IEC 61000-4-4:2012
± 2 kV
100 kHz repetition frequency
Compliant
Immunity to interference, surges
IEC 61000-4-5:2005
± 0.5 kV, ± 1 kV
Compliant
Immunity to interference, line-conducted
disturbances induced by high-frequency fields –
AC voltage grid
IEC 61000-4-6:2013
3 V
0.15 - 80 MHz
6 V
ISM frequency bands
0.15 - 80 MHz
80 % AM at 1 kHz
Compliant
Immunity to interference due to voltage dips,
short interruptions and voltage variations
IEC 61000-4-11:2004
Compliant
Electromagnetic compatibility (EMC) Interference immunity measurements
SIP/SOP
Immunity to interference, discharge of static
electricity
IEC 61000-4-2:2008
± 8 kV contact
± 2kV, ± 4 kV, ± 8 kV, ± 15 kV air
Compliant
Immunity to interference, rapid transient bursts –
I/O, SIP/SOP ports
IEC 61000-4-4:2012
± 1 kV
100 kHz repetition frequency
Compliant