80 PI-MAX
®
3 System Manual Version 1.A
Dual Trigger Mode
Figure 56. DIF Operation: Dual Trigger Timing Diagram
1. The first requirement is that the PI-MAX3
camera be aligned and focused on the area
of interest in the experiment. This is best
accomplished while the PI-MAX3 is
operating in Interline mode (i.e., before
switching to DIF mode). Verify that the
Phosphor Decay Time is appropriate to
the phosphor used by your camera: (this
field is located on the Setup| Hardware
Setup|Controller/ Camera tab). The
procedure for initial focus is outlined in
Chapter 4.
2. After the alignment and focus, the
PI-MAX3 system needs to be put into DIF
mode. On the Setup|Hardware
Setup|Controller/Camera tab, select
Dual Image Feature as the Readout
Mode, and then click OK.
Figure 57. Hardware
Setup|Controller/Camera tab