Cardiac Stress Treadmills Service Manual Preventative Maintenance 5-3
2. Vacuum the areas behind the AC module (deck side). Do not vacuum
the drive board on the AC Drive Module.
3. Vacuum the fan opening.
4. Replace the hood.
Replacement Schedules
Belts should be replaced if they are frayed or show excessive wear.
Electrical Testing
Electrical testing is to be done by the facility’s biomedical department as
required. Check the leakage current of the treadmill periodically—at least
every nine months—to be sure it does not exceed local standards. See
Appendix B, Specifications, for limitations.