EasyManuals Logo

RADIOMETER ABL800 FLEX Reference Manual

RADIOMETER ABL800 FLEX
206 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #10 background imageLoading...
Page #10 background image
ABL800 FLEX Reference Manual 1. Potentiometric measuring principles
General information, Continued
Calibration line
(continued)
To describe the actual condition of the electrode, its calibration line is compared to
the calibration line of the theoretical electrode.
Theoretical calibration line
pH
Measured
potential
(mV)
112.4
75.5
6.8
7.4
The theoretical electrode is
defined to measure the
following:
112.4 mV at pH 7.400,
75.5 mV at pH 6.800.
The position and slope of the calibration line compared to the theoretical
calibration line are described by the status and sensitivity.
Sensitivity
The electrode sensitivity illustrates the slope of the calibration line compared to the
slope of the theoretical electrode.
The sensitivity of the theoretical electrode is 100 % or 1.00.
Theoretical calibration line
Slope= 61.5 mV/pH
Sensitivity = 100 %
pH
Measured
potential
(mV)
112.4
75.5
6.8
7.4
2-point calibration line
Slope =
58.4 mV/pH
Sensitivity = 95 %
If an electrode
has a sensitivity
of 95 % or 0.95,
its sensitivity is
5 % lower than
the theoretical
electrode.
The sensitivity of an electrode is calculated as:
(%)
)802.6398.7(5.61
398.7802.6
×
=
atPotentialatPotential
ySensitivit
where 61.5 = sensitivity of theoretical electrode.
Each electrode has its own sensitivity limits.
Continued on next page
1-5

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL800 FLEX and is the answer not in the manual?

RADIOMETER ABL800 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL800 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals