5. Performance characteristics ABL800 FLEX Reference Manual
Performance test results – cK
+
Primary
reference
methods
NIST certified Standard Reference Material SRM 909b (human serum).
Bias
REF
The FLEXMODE on the ABL805/35 analyzers was tested using SRM 909b:
cK
+
(mmol/L) Bias
REF
N
3.424 -0.03 20
6.278 0.23 20
N = number of measurements on several analysers used for the test.
Bias
ABL
– blood
samples
This bias is presented by the following chart:
Bias (%)