EasyManua.ls Logo

RADIOMETER ABL800 FLEX - Units and Numerical Format of Derived Parameters

RADIOMETER ABL800 FLEX
206 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6. Parameters ABL800 FLEX Reference Manual
Units and numerical format of derived parameters
Calculated
versus estimated
parameters
Derived parameters are calculated or estimated on the basis of measured and keyed
in data. Calculations are made using equations programmed into the analyzer. The
accuracy of the calculations depends on the input parameters keyed into the
analyzer’s computer.
If the calculation of a parameter requires input from the operator, but this input is
not forthcoming, the analyzer will use certain default values (refer to the section
Default Values in this chapter).
Not all input parameters are stored as defaults. In these instances the dependent
derived parameter will not be reported if the relevant input parameter(s) is/are not
entered.
If the default values are used in the calculation of a parameter, then a parameter is
considered estimated (“e”) rather than calculated (“c”).
Acid-base
parameters
The table below lists the acid-base derived parameters.
(ABL83X FLEX corresponds to ABL82X FLEX, but it can measure ctBil and
FHbF).
Symbol Unit Numerical
format of
result
ABL
805
FLEX
ABL8
10/15/
20
FLEX
ABL820/
25/30/35
FLEX
Input
parameter
Sample
type
pH(T) - x.xxx c c c
T
cH
+
(T) nmol/L xxx.x c c c
T
pCO
2
(T) mmHg; torr
kPa
xxx.x
xx.xx
c
c
c
c
c
c
T
cHCO
3
(P) mmol/L xx.x c c c
cBase(B) mmol/L Range: ±30.0 c c c ctHb
e c c
cBase(B,ox) mmol/L xxx.x e c c ctHb
e c c
cBase(Ecf) mmol/L Range: ±30.0 c c c
cBase(Ecf,ox) mmol/L xxx.x e c c
cHCO
3
(P,st) mmol/L xx.x c c c ctHb
e c c
ctCO
2
(P)
Vol %,
mL/dL,
mmol/L
xx.x c
c
c
Continued on next page
6-22

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Related product manuals