4. User-defined corrections ABL800 FLEX Reference Manual
Correction factors for oximetry parameters and bilirubin
Allowed
corrections
The following corrections can be user-defined for the oximetry parameters and
bilirubin:
Parameter Allowed User-defined Corrections
Slope Offset
ctHb
Yes No
sO
2
Yes Yes
FCOHb
No Yes
FMetHb
No Yes
FO Hb
2
No No
FHHb
No No
FHbF
Yes Yes
ctBil
Yes Yes
NOTE: In order to define the corrections accurately, the measurements of the
oximetry parameters and bilirubin on the ABL800 FLEX analyzers should be made
without any entered corrections. To avoid truncation errors from an enabled “Out
of range suppression” function it is important to disable the function.
The following recommendations apply to ctHb:
ctHb
Item Description
Units g/dL; g/L; mmol/L
Sample
Set
ctHb of a SAT100 sample to 15 g/dL (9.3 mmol/L) and
pH 7.4
ctHb, maximum
point
Uncorrected or corrected: 15 g/dL or 9.3 mmol/L
Slope 0.950 - 1.050
The following recommendations apply to sO
2
:
sO
2
Item Description
Units Fraction
Sample
Set
ctHb of gas equilibrated SAT0 and SAT100 samples to
15 g/dL (9.3 mmol/L) and pH 7.4
Slope 0.900 - 1.100
Offset
± 0.050
Continued on next page
4-4

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL800 FLEX and is the answer not in the manual?

RADIOMETER ABL800 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL800 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals