5. Performance characteristics ABL800 FLEX Reference Manual
Performance test results – cGlu
Primary
reference
method
Spectrophotometry, using the hexokinase (HK) method recommended by NCCLS
[5], measured on serum.
Bias
REF
The FLEXMODE on the ABL805/35 analyzers was tested:
cGlu(mmol/L) Bias
REF
N
0.08 0.03 30
2.09 0.06 30
5.08 0.12 30
14.73 -0.02 30
N = number of measurements on several analysers used for the test.
Bias
ABL
– blood
samples
This bias is presented by the following chart:
Bias (%)
A
BL735 (mmol/L)
Continued on next page
5-24

Table of Contents