5. Performance characteristics ABL800 FLEX Reference Manual
Performance test results – ctHb
Primary
reference
method
HiCN method recommended by NCCLS [6].
Bias
REF
The FLEXMODE on the ABL830/35 analyzers was tested:
ctHb (mmol/L) Bias
REF
N
15 (SAT0) 0.32 145
7 (SAT100) 0.04 145
15 (SAT100) 0.37 145
25 (SAT100) 0.97 145
N = number of measurements on several analysers used for the test.
Bias
ABL
This bias is presented by the following chart:
Bias (%)