5. Performance characteristics ABL800 FLEX Reference Manual
Performance test results – cLac
Primary
reference
methods
Spectrophotometry using a lactate dehydrogenase (LDH) method, measured on
serum [10].
Bias
REF
The FLEXMODE on the ABL805/35 analyzers was tested:
cLac (mmol/L) Bias
REF
N
0.36 -0.08 30
2.06 0.12 30
8.3 -0.81 30
11.3 -0.62 30
N = number of measurements on several analysers used for the test.
This bias is presented by the following chart:
Bias
ABL
– blood
samples
Bias (%)
A
BL735 (mmol/L)
Continued on next page
5-26