ABL800 FLEX Reference Manual 7. Solutions and gas mixtures
Electrolyte solutions
List of solutions
The following electrolyte solutions contained in the electrode jackets of the
Radiometer electrodes are used:
Electrolyte for… Quantity Composition
E1001 reference
electrode
0.6 mL in 4 pre-filled
electrode jackets per D711
Membrane Box
Organic compounds and
inorganic salts*
E788 pCO
2
electrode
0.6 mL in 4 pre-filled
electrode jackets per D788
Membrane Box
Inorganic salts, buffer,
hygroscopic compound,
preservative and surfactant.
E799 pO
2
electrode
0.6 mL in 4 pre-filled
electrode jackets per D799
Membrane Box
Inorganic salts, organic
compounds, buffer,
preservative and surfactant.
E722 K electrode 0.6 mL in 4 pre-filled
electrode jackets per D722
Membrane Box
Organic compounds,
inorganic salts, buffer, acid,
and preservative.
E755 Na electrode 0.6 mL in 4 pre-filled
electrode jackets per D755
Membrane Box
Inorganic salts, organic
compounds, preservative and
surfactant.
E733 Ca electrode 0.6 mL in 4 pre-filled
electrode jackets per D733
Membrane Box
Inorganic salts, organic
compounds, buffer,
preservative and surfactant.
E744 Cl electrode 0.6 mL in 4 pre-filled
electrode jackets per D744
Membrane Box
Inorganic salts, organic
compounds, preservative,
surfactant and hygroscopic
products.
E7066 Glucose
and E7077 Lactate
electrodes
0.6 mL in 5 plastic
capsules to fill the
electrode jackets (4 units)
per D7066 and D7077
Membrane Boxes
Buffer, inorganic salts,
thickening agent, preservative
and surfactant.
*WARNING/CAUTION: Irritating to eyes, respiratory system and skin. In case of
contact with eyes, rinse immediately with plenty of water and seek medical advic
e.
Storage
Temperature: Electrolyte solution:
2-25
o
C (36-77
o
F) For glucose electrode
2-10
o
C (36-50
o
F) For lactate electrode
2-32
o
C (36-90
o
F) For all other electrodes
Expiration date and Lot No. are printed on a label on the side of the membrane
box.
Stability
- 7-5