5. Performance characteristics ABL800 FLEX Reference Manual
Performance test results – cCl
Primary
reference
method
NIST certified Standard Reference Material SRM 909b (human serum).
Bias
REF
The FLEXMODE on the ABL805/35 analyzers was tested using SRM 909b:
cCl
(mmol/L) Bias
REF
N
89.11 0.6 20
119.43 2.4 20
N = number of measurements on several analysers used for the test.
Bias
ABL
– blood
samples
This bias is presented by the following chart:
Bias
(
%
)
A
BL735 (mmol/L)
Continued on next page