EasyManuals Logo

RADIOMETER ABL800 FLEX Reference Manual

RADIOMETER ABL800 FLEX
206 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #104 background imageLoading...
Page #104 background image
5. Performance characteristics ABL800 FLEX Reference Manual
Performance test results – cCa
2+
Primary
reference
methods
The calcium transfer standards were used. These are traceable to NIST SRM915
and have an ionic strength of 160.0 mmol per kg of water and pH 7.40 at 37 °C,
using 1 mmol/L (37 °C) HEPES buffer.
The standards were produced as indicated in [4].
Bias
REF
The FLEXMODE on the ABL805/35 analyzers was tested:
cCa
2+
(mmol/L) Bias
REF
N
0.4879 0.038 150
1.2700 0.025 150
2.5657 0.052 150
N = number of measurements on several analysers used for the test.
Bias
ABL
– blood
samples
This bias is presented by the following chart:
Bias (%)
A
BL735 (mmol/L)
Continued on next page
5-22

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL800 FLEX and is the answer not in the manual?

RADIOMETER ABL800 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL800 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals