ABL800 FLEX Reference Manual 6. Parameters
Conversion of units, Continued
V
·
O
2
V
·
O
2
(mmol/L)/min = V
·
O
2
/22.41 (mL/dL)/min
cGlucose
[22]
cGlucose (mg/dL) =
18.016 × cGlucose (mmol/L) or
cGlucose (mmol/L) =
0.055506 × cGlucose (mg/dL)
cLactate
[22]
cLactate (mg/dL) =
9.008 × cLactate (mmol/L) or
cLactate (mmol/L) =
0.11101 × cLactate (mg/dL)
cLactate (meq/L) =
cLactate (mmol/L)
(conversion based on the molecular weight of lactic acid)
ctBil
ctBil (µmol/L)
=
17.1 × ctBil (mg/dL)
ctBil (µmol/L)
=
1.71 × ctBil (mg/L) or
ctBil (mg/dL) =
0.0585 × ctBil (µmol/L)
ctBil (mg/L) =
0.585 × ctBil (µmol/L)
NOTE:
All conversions of units are made by the analyzer.
6-49