ABL800 FLEX Reference Manual 4. User-defined corrections
General information, Continued
The slope/offset for each parameter are configured in the Parameters Setup under
General Setup. User-corrected values are marked with a “*” after the result.
Entering
user-defined
corrections
NOTE: The user-defined corrections will be applied to measurements on QC
solution unless the "Apply parameter corrections to QC" option was
deactivated in Miscellaneous Setup.
For detailed instructions on how to enter user-defined corrections, refer to the
section
Parameter Setup in Chapter 3 of the Operator’s Manual.
4-3