ABL800 FLEX Reference Manual 4. User-defined corrections
Correction factors for oximetry parameters and bilirubin,
Continued
FCOHb
The following recommendations apply to FCOHb:
Item Description
Units Fraction
Sample
The zero point (
FCOHb 0) is saturated to approximately
SAT100, and
ctHb is set to 15 g/dL (9.3 mmol/L) and pH
7.4.
Offset
± 0.050
The following recommendations apply to FMetHb:
FMetHb
Item Description
Units Fraction
Sample
The zero point (
FMetHb 0) is saturated to approximately
SAT100, and
ctHb is set to 15 g/dL (9.3 mmol/L) and pH
7.4.
Offset
± 0.050
The following recommendations apply to FHbF:
FHbF
Item Description
Units Fraction
Sample Radiometer recommends that ctHb in the adult samples (with
FHbF = 0) and fetal samples (with high FHbF) is set to 15
g/dL (9.3 mmol/L), sO
2
100 %, and pH 7.4.
The “Correction for HbF levels less than 20 %” function
should be enabled in order to have the FHbF value displayed
for the adult sample.
Averaging repeated measurements on blood from different
donors gives an optimized accuracy of the correction.
Averaging repeated measurements on blood from the same
donor also improves the accuracy.
Slope 0.800 - 1.200
Offset
± 0.20
Continued on next page
4-5

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL800 FLEX and is the answer not in the manual?

RADIOMETER ABL800 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL800 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals