ABL800 FLEX Reference Manual 5. Performance characteristics
Performance test results – pO
2
, Continued
Bias and
imprecision –
expired air
samples
The bias and imprecision for expired air samples are as follows:
pO
2
, mmHg Bias
ABL835/30/25/20/15/10/05
15 0.8
40 0.4
130
0.4
230
0.9
570 4.2
Imprecision:
pO
2
mmHg S
0
S
D
S
ABL
S
X
15 0.3 0.3 1.2 1.3
40 0.3 0.3 1.0 1.1
130 0.3 0.3 0.7 0.8
230 2 2 3 4
570 5 5 13 15
15 ABL700 Series and ABL800 FLEX analyzers are tested over three days for all
levels. Bias is determined against certified gases at sea level.
5-15