5. Performance characteristics ABL800 FLEX Reference Manual
Interference tests, Continued
ctBil sensitivity
for MCHC
variations
(continued)
c
c
tBil
tBil
0.58
1 0 350
=−
×
=+
.
.
58
18
0071 And ctBil = 0.071 x 400 = 28 µmol/L.
If the reference value for Hct is known, it is possible to correct the displayed ctBil
value, using the following equation:
cc
c
tBil(corrected) tBil(displayed
tHb(displayed)
Hct(reference)
−×
)
.1 00301
1
ctHb is measured in g/dL.
ctBil is slightly sensitive to pH deviations from the nominal value of pH = 7.4.
ctBil sensitivity
for pH changes
The following table shows the changes in ctBil compared to the value at pH =
7.4.
Sample Type
ctHb
g/dL
Nominal
ctBil
µmol/L
ctBil
(7.47.1)
µmol/L
ctBil
(7.47.9)
µmol/L
Adult/fetal plasma 0 0 3 0
Adult blood, sO
2
= 100 % 15 0 0
5
Fetal blood, sO
2
= 100 % 15 0
13
4
Adult/fetal plasma spiked with
unconjugated bilirubin
0 400
2 1
Adult/fetal plasma spiked with
conjugated bilirubin
0 400 9
11
Adult blood spiked with un-
conjugated bilirubin, sO
2
= 100 %
15 400 10
26
Fetal blood spiked with un-
conjugated bilirubin, sO
2
= 100 %
15 400
4 16
Adult blood spiked with conjugated
bilirubin, sO
2
= 100 %
15 400 14
35
Fetal blood spiked with conjugated
bilirubin, sO
2
= 100 %
15 400 0
26
5-54