ABL800 FLEX Reference Manual 6. Parameters
Measured parameters, Continued
ctHb
Definition Concentration of total hemoglobin in blood.
Total hemoglobin includes all types of hemoglobin:
deoxy-, oxy-, carboxy-, met-.
Unit g/dL; g/L; mmol/L
Measuring range g/dL: 0.00- 27.7
g/L: 0.0-277
mmol/L: 0.00-17.2
Reference range g/dL: 13.5-17.5 (m); 12.0-16.0 (f)
g/L: 135-175 (m); 120-160 (f)
mmol/L: 8.4-10.9 (m); 7.4-9.9 (f)
Conversion of units
ctHb (g/dL) = 1.61140
× ctHb (mmol/L);
ctHb (g/L) = 16.1140
× ctHb (mmol/L);
ctHb (mmol/L) = 0.62058
× ctHb (g/dL) =
0.062058
× ctHb (g/L)
Default value: 9.3087 mmol/L, (15.0 g/dL or 150 g/L)
sO
2
Can also be calculated.
Definition Oxygen saturation, the ratio between the concentrations
of oxyhemoglobin and the hemoglobin minus the
dyshemoglobins.
Unit %; fraction
Measuring range %: 0.0-100.0
Fraction: 0.000-1.000
Reference range %: 95-99 (m, f)
Fraction: 0.95-0.99 (m, f)
Continued on next page