6. Parameters ABL800 FLEX Reference Manual
Measured parameters, Continued
FHHb
(continued)
Equation
FHHb = 1- sO
2
× (1FCOHbFMetHb)FCOHb
FMetHb
If sO
2
is not measured, it will be calculated from
equation 39.
If dyshemoglobins (FCOHb, FMetHb) are not known,
they are set to the default values.
FHbF
Definition Fraction of fetal hemoglobin in total hemoglobin in
blood
Unit %; fraction
Measuring range %: 0-100
Fraction: 0.00-1.00
Reference range
(neonates)
%: 80 (m, f)
Fraction: 0.80 (m, f)
cK
+
Definition Concentration of potassium ions in plasma.
Unit mmol/L; meq/L
Measuring range mmol/L; meq/L: 0.5-25.0
Reference range m, f: 3.4 – 4.5 mmol/L
Conversion of units mmol/L = meq/L
cNa
+
Definition Concentration of sodium ions in plasma.
Unit mmol/L; meq/L
Measuring range mmol/L; meq/L: 7-350
Reference range m, f; 136 – 146 mmol/L
Conversion of units mmol/L = meq/L