EasyManuals Logo

RADIOMETER ABL800 FLEX Reference Manual

RADIOMETER ABL800 FLEX
206 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #152 background imageLoading...
Page #152 background image
ABL800 FLEX Reference Manual 6. Parameters
Input parameters, Continued
pO
2
(v
)
Definition Oxygen tension of mixed venous blood.
Input range/Unit mmHg; torr: 0.0-750.0
kPa: 0.00-100
Conversion
p(kPa) =0.133322 × p(mmHg)
p(mmHg) = (kPa)500638.7 p
×
sO
2
(v
)
Definition Oxygen saturation of mixed venous blood.
Input range/Unit %: 0.0 100.0
fraction: 0.000 – 1.000
Q
·
t
Definition Cardiac output; volume of blood delivered from
the left ventricle into the aorta per unit of time.
Also termed CO or C.O.
Input range/Unit 0.0 - 1000.0 L/min
V
·
O
2
Definition Oxygen consumption; total amount of oxygen
utilized by the whole organism per unit of time.
Input range/Unit mL/min: 0 - xxxx
mmol/min: 0.0 - xxx.x
Conversion (mmol/L)min = (mL/dLmin)/22.41
VCO
Definition Volume of carbon monoxide added to the patient
for measurement and calculation of V(B) [5].
Input range/Unit 0.0 - 1000.0 mL
Fortsættes på næste side
6-15

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL800 FLEX and is the answer not in the manual?

RADIOMETER ABL800 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL800 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals