6. Parameters ABL800 FLEX Reference Manual
Units and numerical format of derived parameters, Continued
(ABL83X FLEX corresponds to ABL82X FLEX, but it can measure ctBil and
FHbF).
Oxygen
parameters
(continued)
Symbol Unit Numerical
format of
result
ABL
805
FLEX
ABL
810/15
/20
FLEX
ABL
820/25
/30/35
FLEX
Input
parameter
Sample
type
RI(T) % xx e e c FO
2
(I)+RQ+T Arterial,
fraction x.xx e e e
T
capillary
VO
2
/V(dry air) % xxx.x c c c
fraction x.xxx
Q
x
- xx.x e e* c* ctHb
1)
+p50(st)
1)
Arterial,
e e* c* capillary
V(B) L x.x c c c ctHb+VCO+FCOH
b(1)+FCOHb(2)
* If the sO
2
value for establishing the ODC is greater than 0.97, the calculation of
the parameter is not performed unless the p50(st) value is keyed in.
1)
If not measured, e.g. ctHb (or derived by analyzer, e.g. p50(st)).
Electrolyte
Parameters
The table below lists the electrolyte derived parameters for the ABL800 FLEX
analyzers.
Symbol Unit Numerical
format of
result
ABL8X5 FLEX Input parameter Sample
type
Anion Gap, K
+
meq/L,
mmol/L
xxx.x c
2)
Anion Gap meq/L,
mmol/L
xxx.x c
3)
cCa
2+
(7.4) meq/L,
mg/dL,
mmol/L
xx.x c
4)
mOsm mmol/kg xxx.x c
5)
2) If the analyzer includes K, Na and Cl electrodes.
3) If the analyzer includes Na and Cl electrodes.
4) If the analyzer includes Ca electrode.
5) If the analyzer includes Na and Glucose electrodes.
6-26

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL800 FLEX and is the answer not in the manual?

RADIOMETER ABL800 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL800 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals