1. Potentiometric measuring principles ABL800 FLEX Reference Manual
pH electrode, Continued
Corrections
(continued)
ABL8XX
FLEX
Mode A
0
A
1
Equation
35/25/15 FLEXMODE (message 870) 1.030
−0.216
A+
(cont.) FLEXMODE (message 869) 1.030
−0.216
A+
30/20/10 S85 0,9964 0.0150 A
C55 1.025
−0.1880
A+
FLEXMODE (no message) 1.0006
−0.0035
A+
FLEXMODE (message 872) 1.0209
−0.1575
A+
FLEXMODE (message 871) 1.025
−0.1880
A+
FLEXMODE (message 870) 1.030
−0.216
A+
FLEXMODE (message 869) 1.030
−0.216
A+
05 S165 0,9964 0.0150 A
S95 0,9964 0.0150 A
S85 0,9964 0.0150 A
C95 1.007
−0.053
A+
C55 1.025
−0.1880
A+
FLEXMODE (no message) 0,9964 0.0150 A+
FLEXMODE (message 874) 1.007
−0.053
A+
FLEXMODE (message 873) 1.007
−0.053
A+
FLEXMODE (message 872) 1.0216
−0.1639
A+
FLEXMODE (message 871) 1.025
−0.1880
A+
FLEXMODE (message 870) 1.030
−0.216
A+
FLEXMODE (message 869) 1.030
−0.216
A+
00
Continued on next page
1-12