1. Potentiometric measuring principles ABL800 FLEX Reference Manual
pCO
2
electrode, Continued
Corrections -
blood samples
(continued)
B = barometric pressure in kPa
pH
2
O = partial pressure of saturated water vapor (6.2751
kPa)
B
0
= correction cut-off
B
1
= correction factor
ABL8XX
FLEX
Mode A
o
A
1
A
2
A
3
B
0
B
1
Eq.
35/25/15 S195 -0.003573 1.1126 0.0051 -0.0000002 A
S95 -0.003573 1.1126 0.0051 -0.0000002 1.000 -0.016 A, B
S85 -0.003573 1.1126 0.0051 -0.0000002 A
C95 -0.003573 1.1126 0.0051 -0.0000002 1.013 0.010 A, B
C55 -0.003573 1.1126 0.0051 -0.0000002 1.12 -0.28 A, B
*FM (no
message)
-0.003573 1.1126 0.0051 -0.0000002 A, B
*FM (message
874)
-0.003573 1.1126 0.0051 -0.0000002 1.013 0.010 A, B
*FM (message
873)
-0.003573 1.1126 0.0051 -0.0000002 1.013 0.010 A, B
*FM (message
872)
-0.003573 1.1126 0.0051 -0.0000002 1.0884 -0.1619 A, B
*FM (message
871)
-0.003573 1.1126 0.0051 -0.0000002 1.090 -0.150 A, B
30/20/10 S85 -0.003573 1.1126 0.0051 -0.0000002 A
C 55 -0.003573 1.1126 0.0051 -0.0000002 1.12 -0.28 A, B
*FM (no
message)
-0.003573 1.1126 0.0051 -0.0000002 1.013 0.050 A, B
*FM (message
872)
-0.003573 1.1126 0.0051 -0.0000002 1.0819 -0.0495 A, B
*FM (message
871)
-0.003573 1.1126 0.0051 -0.0000002 1.090 -0.150 A, B
*FM = FLEXMODE
Continued on next page
1-18