EasyManua.ls Logo

RADIOMETER ABL800 FLEX - Page 55

RADIOMETER ABL800 FLEX
206 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2. Amperometric measuring principles ABL800 FLEX Reference Manual
Metabolite electrodes, Continued
where:
Sensitivity
(continued)
cX(Cal 1) = Actual concentration of glucose/lactate in the Cal 1
solution.
I
0
(final) = Extrapolated final zero current value of the metabolite
electrode at the time of the last updating.
TI(Cal 1)
= electrode current due to presence of glucose/lactate.
The sensitivity limits of the metabolite electrodes are as follows:
Electrode Sensitivity Limits
Glucose 100 - 1800 pA/mM
Lactate 150 - 2000 pA/mM
Drift
The drift in the sensitivity of the metabolite electrodes is calculated from the
following equations:
1)X(Cal
Sens
(final)Ifinal)1,I(Cal
Drift
0
c
=
where:
I(Cal 1,final) = Current at the final measurement on Cal 1 solution.
Sens = Sensitivity of the glucose/lactate electrode from the previous
calibration.
cX(Cal 1) = Actual concentration of glucose/lactate in the Cal 1 solution.
I
0
(final) = Extrapolated final zero current value of the metabolite
electrode measured at the time of the last updating.
The default drift tolerances set by Radiometer for the metabolite electrodes are:
± 0.5 mM for the glucose electrode
± 0.2 mM for the lactate electrode.
Continued on next page
2-16

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Related product manuals