2. Amperometric measuring principles ABL800 FLEX Reference Manual
Metabolite electrodes, Continued
Corrections
(continued)
Corrections for cGlu :
ABL8XX
FLEX
Mode A
0
A
1
Equation
35/25/15 S195 0.94 0.1 A
S95 1.00 0.0 A, B
C95 1.06 0.0 A, B
C35 1.16 0.0 A, B
*FM (no message) 0.94 0.1 A
*FM (message 874) 1.06 0.0 A, B
*FM (message 873) 1.06 0.0 A, B
05 S165 0.94 0.1 A
S95 1.00 0.0 A, B
C95 1.06 0.0 A, B
C35 1.16 0.0 A, B
*FM (no message) 0.94 0.1 A
*FM (message 874) 1.06 0.0 A, B
*FM (message 873) 1.06 0.0 A, B
*FM = FLEXMODE.
Corrections for cLac:
ABL8XX
FLEX
Mode A
0
A
1
Equation
35/25/15 S195 0.97 -0.04 A
S95 1.03 0.03 A, B
C95 1.03 0.18 A, B
C35 1.13 0.05 A, B
*FM (no message) 0.97 -0.04 A
*FM (message 874) 1.03 0.18 A, B
*FM (message 873) 1.03 0.18 A, B
*FM = FLEXMODE.
Continued on next page
2-18