The time delay enables the user to take the weighed load of the weighing pan, if a
weighing process is in progress. Pressing T/O key causes temporary delay of the
adjustment process initiation.
Automatic adjustment settings
Fig. 12. Settings of automatic balance adjustment
1 main menu number
2 marker of selected function
3 function name
4 name of an active function / mode / process
5 selecting factor triggering auto-adjustment process (time / temperature)
6 determining time interval between the following auto-adjustment
processes
7 value of set auto-adjustment triggering factors
8 value of set time interval between the following automatic adjustment
processes
Changing the values of automatic adjustment triggering factor and automatic adjustment
time causes changes in description of the above fields (fields in fig no. 9 and 10).
01 Internal automatic adjustment
Initiates internal automatic adjustment process, which is carried out fully automatically
with no operator’s activity. If balance’s weighing pan is loaded, then the balance
displays a command ordering unloading it.
02 External adjustment
Adjustment process carried out with an external weight, which value is saved in
balance’s memory. The function is disabled in verified balances.
03 User adjustment
Adjustment process carried out with an optional weight. Mass of the weight is specified
before process initiation. The function is disabled in verified balances.

Table of Contents

Other manuals for RADWAG AS series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS series and is the answer not in the manual?

RADWAG AS series Specifications

General IconGeneral
DisplayLCD with backlight
Maximum CapacityVaries by model
ReadabilityVaries by model
Pan SizeVaries by model
CalibrationInternal or external, depending on model
Stabilization Time≤ 3 seconds
InterfaceRS232, USB
Repeatability (Standard deviation)Varies by model
LinearityVaries by model
Net WeightVaries by model
Protection ClassIP 43