7. The density of air is calculated automatically.
Fig. 23-9. Air buoyancy correction the air density correcting coefficient
At this stage the user can:
Restart the procedure from the beginning (by pressing Units key)
Return to weighing without saving changes on air density determination
(press MODE key and select mode WEIGHING)
Accept calculated density value.
Fig. 23-10. Air buoyancy correction Message window
8. The display indicates the calculated values the balance is ready for operation
with the determined air density coefficient.
9. Return to weighing by selecting WEIGHING mode.
10. Set the attribute of air buoyancy correction to YES